Anaerobic Jar Maintenance
For the stainless steel jars, common detergent and disinfectant is recommended to use for the jar cleaning, and then rinse the jar with clean water.
While for the PMMA jars,
1、Never to be cleaned with alcohol
2、Not to be exposed to temperatures more than 80℃
3、Never touch the jar with anything sharp and do not drop the jar
4、Neutral detergent and disinfectant is recommended to use for the jar cleaning, then rinse the jar with clean water.
5、Clean and apply acid-free vaseline or silicon oil on the rubber O-ring of the lid, you can do this using a swab or finger